Shutting Down - Emptying The Humidifier - Thermo Scientific Vitrobot Mk IV User Manual

Table of Contents

Advertisement

Chapter | Operating Instructions
3.11 Shutting down – Emptying the humidifier
After shutting down the Vitrobot PC, the remaining water in the humidifier may be removed.
To do this, pull the metal ring connector downwards to disconnect the electronic cable from
the humidifier. Subsequently, twist and pull the humidifier in order to unleash the
bajonet-connection. Emptying the humidifier is a two step process:
Pouring the humidifier will empty the central reservoir.
1.
Re-connecting the syringe to the plastic tube at the bottom part and removing the water
2.
from the outside reservoir is the second essential step
See the video movie for more details.
Note
To prevent any formation of bacteria in the water of the humidifier, it is advisable to
dispose the water supply in the humidifier at the end of each working day!
Confidential, limited rights
PN 103261 | Revision D | 23-Oct-2019
User Manual
Page 31