Merck Millipore Mobius Chrom 20 System User Manual page 71

With flexware assemblies
Table of Contents

Advertisement

Mobius
Chrom 20 System with Flexware
®
Running the Operation
To run the operation downloaded to the PLC, exit the recipe editor by clicking on the Exit icon in
the Recipe Editor toolbar and return to the Process Display.
Note
Always check the Process Display for proper system status before running an operation.
Ensure that no manual forces have been applied to valves, as these will override any
operation criteria that are specified. The proper setting for these is Auto.
In the Process Display, click on the Run icon in the tool bar. This causes the Run Header Data
Form to display and allows the user to enter the Run Header data.
This form allows the operator to enter run header data to identify the run. Once again, the
rows in white can be edited. The CCP
unique run ID. By default, the software supplies a unique run ID by providing a date and time
stamp. The user can replace this ID with her or his own unique identifier. For more details on
this form, refer to the chapter on Batch Reporting.
Clicking on the OK button on the Run Header Data form closes it, returns the operator to the
Process Display and begins execution of the operation stored in the PLC.
Managing Procedures
Creating a Procedure
To create a procedure, click on Procedure and then Edit Procedure in the Recipe Editor Menu bar
or click the icon in the tool bar.
The Edit Procedure window opens.
In the lower right corner of the form, the Operation Files list box presents all the operations
currently stored within the path indicated in the Operation Files Path list box. To add an
Operation to the Procedure, select the Operation from the list of Operations on the right and
click the Add Operation Button. The Operation is added to the Operations in the Procedure list
box located in the lower left corner of the form. To remove operations from this list, click on
Remove Operation.
Note
If the Audit Trail is activated, only validated operations will be present in the Operation
Files list box.
To edit the Procedure Header, click on the Header button. Like an operation, the Last Saved By
field must be filled to save and download the procedure.
Assemblies
®
software batch reporting utility identifies each run with a
®
71