Manifold Assembly Onto The Manifold - Merck Millipore Mobius Chrom 20 System User Manual

With flexware assemblies
Table of Contents

Advertisement

130
Installing Mobius
Assemblies for Chromatography Pump 2 Top
3
4
5

Manifold Assembly onto the Manifold

XM4P2MTC
The Pump Manifold Assemblies are available with TC fittings. All fittings are shipped with a cap
installed. Remove the covering just before connecting the flow path to the system.
Pump Manifold Assembly XM4P2MTC
1. Unpack the manifold assembly
Manifold
2. Using the touch screen, open valves XV007, XV008, XV009, XV0010, XV0011, or use the
Open All Valves button.
3. Install the tubing into the opened pinch valves on the manifold from the right to the left.
4. Using the touch screen close the pinch valves XV007, XV008, XV009, XV0010, XV0011 or
use the Close All Valves button.
®
6
7
THIS DRAWING IS
THE PROPERTY OF
MILLIPORESIGMA
00013746FM VER3.0
2B
2C
2D
2E
2A
1B
1C
1D
1E
1A
Chrom 20 with Flexware
8
9
10
Connect
TO P002
This drawing and the data contained thereon is confidential
and the property of MilliporeSigma, Billerica Massachusetts,
USA. Third Parties may use this document only to review
and approve the drawing for use by MilliporeSigma in its
manufacturing. Any other use or reproduction of this
drawing is expressly prohibited without MilliporeSigma's
written consent.
TOLERANCE BLOCK
DRAWING NUMBER:
XM4P2MTC
X
± 1
X.X
± 0.5
DESCRIPTION:
MOBIUS® ASSEMBLY
X.XX
± 0.25
X.XXX
± 0.125
± 5°
DRAWN BY:
A. CHASSE
ALL DIMENSIONS IN INCHES
11
12
USER GUIDE VIEWS
REV:
A
SHEET SIZE: SCALE:
SHEET:
AUTHORING SITE:
DATE:
B
NTS
MM-DANVERS
2 of 14
17-SET-2019