Mobius
Chrom 20 System with Flexware
®
Process Controls Bar
The content of the Process Controls Bar varies based on the current system status (alarm
presence, recipe running, ...).
Symbol
Assemblies
®
START
Starts the current recipe.
ABORT
Aborts the current recipe.
HOLD
Sets the system to hold state.
RESUME
Resumes from hold state.
PAUSE
Pauses the currently running recipe.
RESUME
Resumes the running recipe from pause.
Opens the recipe step jump popup.
Visible when an operation or procedure is being executed. A
jump step can occur to any valid step in the same phase, but
not to a different phase.
Opens the Alarm history popup.
Acknowledges current alarms.
If at least one unacknowledged critical alarm is present, then
the button appears in red.
Stops the alarm buzzer.
Creates a screenshot and save it as a jpg file.
The default file location is:
D:\Millipore\PrintScreens
Filename is automatically created based on the current date
& time.
Records a comment within the System events.
Trends opens the Trend chart panel.
Maximize Switch from the trend chart panel to the full screen
chart.
Minimize Switch from the full chart window to the panel chart.
Opens alarm status display.
Opens display setup popup.