Changing The Valve Pads - Merck Millipore Mobius Chrom 20 System User Manual

With flexware assemblies
Table of Contents

Advertisement

202

Changing the Valve Pads

The valve pads in the Flexware
after six months of use.
1. Run the Unlock Door recipe to open the door and access the valve pads.
2. For Valve Pads A, B and C, insert the blade of a flat head screwdriver between the valve
pads and the Flexware
Clamshell Assembly.
3. For Valve Pad D, pull the valve pad out using the stems.
4. Discard the used pads.
5. Install the new pads by aligning the pins on the pads with the holes in the Flexware
Clamshell Assembly. Push the pins into the holes to secure the pads.
6. Run the Lock Door recipe to close secure the door.
D
Valve Pad Locations
Clamshell Assembly should be changed after 2500 cycles or
®
Clamshell Assembly and pry the valve pads off of the Flexware
®
D
D
A
D
A
D
D
D
D
D
D
D
D
D
A
B
C
C
C
D
C
User Guide
®
®