Section 4
2. Use the numeric keys to type the Network ID number.
3. Press <ENT> to store Network ID. This moves the cursor to [RTS Key Time].

RTS Key Time

During a serial transmission session, you can select how long the flow computer holds its request-to-
send line high until it receives a clear-to-send signal. If the RTS key time expires, the FUG1010GCN will
return its RTS line low and abort the transmission attempt. Data entry units are seconds. The default RTS
key time is 0.2 seconds.
NOTE: The RTS Key Time function is only provided when the meter's network ID number is non-
zero.
To set the RTS Key Time:
1. Press <Right Arrow> to access the [RTS Key Time] option list.
2. Use the <Up or Down Arrow> to move the asterisk to the required RTS time.
3. Press <ENT to store the data.

4.12.9 BACKLIGHT

This menu cell allows you to specify how long you want to maintain the backlighting for the LCD Graphic
Display. Select [On] for continuous illumination. Press the <Right Arrow> to access the option list and
then scroll to desired time delay (before turn-off). Press <ENT> to store your setting.

4.12.10 SYSTEM INFO

This menu provides general information about the flowmeter.
Version
Reset Date/Time
Op System P/N
Checksum
Code
System Time
4.13

THE FUG1010GCN GRAPHIC DISPLAY SCREENS

The FUG1010GCN Graphic Display Sub-System is a versatile resource which provides visual access to
all system variables and conditions. Depending upon your selection, details of the display my vary, but
all elements are always clearly labeled with units and variable names. You're always free to move
through the displays by using the arrow keys. The meter will hold the last display selected indefinitely or
until the menu is selected by pressing the <Menu> key. When you again leave the menu, the last se-
lected display will be shown upon your return.
After completing the Site Setup, Check your local display screen to ensure that everything is operating
properly. The flowmeter begins to measure flow as soon as you complete the transducer installation.

SYSTEM INFO MENU STRUCTURE

This is the meter's operating system version number. The Technical Services Group
may request this number during consultations.
Shows the date and time of this meter's last <F4> reset.
Meter's operating system part number.
Operating System verifying code. The Technical Services Group may request
this number during consultations.
Software Compile Date/Time. Operating System identifier. The Technical Ser-
vices Group may request this number during consultations.
Use to set system time and date. Format: xx.xx.xx. xx.xx.xx
4-90
1010GCNFM-3C