HOW TO RECALL A SITE SETUP

The [Recall Site] command allows you to reinstall the system at a former site. For an original installation,
see Create/Name Site for instructions on how to create a new Site Setup. This menu cell provides a list of
saved site names. Scrolling to a site name and pressing <ENT> moves all the parameters associated
with that name into Active Memory.
If there are no site setups present in site storage memory, the menu cell reports [No Sites] in the right-
hand column. Saving a site setup with a site name adds the name to this option list. The first selection on
the list is the currently active Site Setup. The system will be ready for operation after you install the
transducers and make the required hardware connections. Note that the installed transducers must
comply with the recalled site parameters.
To recall a Saved Site Setup:
1. Press the <Right Arrow> to access the list of saved site names.
2. Use the <Down Arrow> to move the cursor to the Site Name.
3. Press <ENT>. This triggers a pop-up window that asks if you intend to use the original transducer
position (recommended). This means that you have to remount the transducers exactly as they were
in the previous install.
4. Press <ENT> to recall the stored site setup. This enables the measurement channel for immediate
operation.
If you cannot recreate the original transducer installation, then press the <Down Arrow> to change the
pop-up list from [Original] to [New]. Press <ENT>.
This disables the measurement channel until you repeat the Pick/Install Spool routine.
4.1.2

HOW TO ENABLE AND DISABLE A MEASUREMENT CHANNEL

The flowmeter disables a measurement channel until the completion of the required Site Setup entries
and the transducer installation procedure. The meter enables the measurement channel automatically
after it receives all required site data and completes the Spool install routine. (The Site Enable menu cell
allows you to disable or enable a measurement channel after there has been a successful transducer
install).
The system does not require your presence for routine operation. If a fault condition (e.g., an empty pipe)
disrupts operation, the system will recover automatically after the fault clears. Use Channel Enable [No]
to turn off a channel that is currently unused but active; for example, if the transducers have been
removed for servicing.
To disable a channel that is operating:
1. Press the <Right Arrow> to access the [Channel Enable] option list.
2. Use the <Up or Down Arrow> to move the cursor to [No].
3. Press <ENT> to disable the measurement channel.
To enable a channel:
1. Press the <Right Arrow> to access the [Channel Enable] option list.
2. Use the <Up or Down Arrows> to move the cursor to [Yes].
3. Press <ENT> to enable the measurement channel.
(Where a visual of the keys is shown; this means press this key)