Section 3
As shown in the steps below, selecting this mode only requires <Right Arrow> key presses. This is be-
cause dual-path is the default configuration for this instrument.
1. Make sure that the keypad enable switch is in the Enable position. Turn on power and press <MENU>
to access the [Meter Type] screen. Refer to the figure below. Note that on the left-hand column [Meter
Type] is highlighted as are all the available meter types on the right-hand column. Press the <Right
Arrow>. This places a cursor (Î) next to [Dual Path Flow].
2. Press the <Right Arrow> again. This selects [Dual Path Flow] menu and then pressing <ENT> will also
access the Dual Path Flow menu.
Creating a New Site Setup (See also paragraph 4.1.3)
The Path Setup section of this manual provides detailed instructions on how to use all the Path Setup
menu commands. This section provides an example of how to use the [Create/Rename Site] command
to create a new site setup. Always begin a new installation by issuing the [Create/Name Site]
command. Although you do not need to enter a site name to create a new site setup, you have to
provide one to save the site setup or to identify the source of logged data. In this example, we will use
[SITE1] for the site name.
Siemens
Dual Path
Select Meter Type
Meter Type
> Dual Path Flow
Meter Facilities
Liquid Class
Siemens
Dual Path
Conforms operation to user Preferences
Channel/Path Setup
Pipe Data
Application Data
Pick/Install Xdcr
Operation Adjust
Flow/Total Units
Data Span/Set/Cal
Stripchart Setup
Datalogger Setup
I/O Data Control
Diagnostic Data
Dual Path Flow
3-8
SITE 1
1010GCNFM-3C