Default value
0.001S
Example
TRAC:TIM <NRf+>
Query command
TRACe:TIMer?
Returns
NR3

TRACe:POINts:ACTual?

This command selects number of actual readings in the cache.
Syntax
TRACe:POINts:ACTual?
Arguments
None
Default value
0
Example
TRAC:POIN:ACT?
Returns
NR1

TRACe:DATA?

This command reads all values saved in the cache.
Syntax
TRACe:DATA?
Note
Before sending the query command TRACe:DATA?, the command TRIGger[:IMMediate]
must be sent to the instrument to trigger the instrument into data storage status. And the
argument of the command TRACe:FEED:CONTrol <NEXT|ALWays|NEVer> must be set
to NEXT or ALWays, otherwise the system will prompt an error.