Emerson Rosemount 2051 Reference Manual page 93

Pressure transmitter with hart revision 5 and 7 selectable protocol
Hide thumbs Also See for Rosemount 2051:
Table of Contents

Advertisement

Reference Manual
00809-0100-4107
5.9.1
Switching HART revision with generic menu
If the HART
device, it should load a generic menu with limited capability. The following procedures
allow for switching between HART Revision 7 and HART Revision 5 from a generic menu.
Procedure
1. Locate "Message" field.
2. To change to HART Revision 5, Enter: HART5 in the message field.
3. To change to HART Revision 7, Enter: HART7 in the message field.
5.9.2
Switching HART revision with Field Communicator
From the HOME screen, enter the Fast Key sequence and follow steps within the Field
Communicator to complete the HART
From the HOME screen, enter the Fast Key
sequence
Device Dashboard Fast Keys
5.9.3
Switching HART revision with AMS Device Manager
Procedure
1. Select Manual Setup and select HART.
2. Select Change HART Revision, then follow the on screen prompts.
5.9.4
Switching HART revision with LOI
Navigate to HART
REV 7. Use
Figure 5-8: Change HART Revision with LOI
Reference Manual
®
configuration tool is not capable of communicating with a HART Revision 7
Note
AMS Device Manager versions 10.5 or greater are compatible with HART Revision 7.
®
REV within the extended menu and select either HART REV 5 or HART
Figure 5-8
below to change HART Revision:
VIEW CONFIG
ZERO TRIM
UNITS
RERANGE
LOOP TEST
DISPLAY
EXTENDED MENU
EXIT MENU
Operation and Maintenance
®
revision change.
HART5
2, 2, 5, 2, 4
EXTENDED MENU
CALIBRAT
DAMPING
TRANSFER FUNCT
SCALED VARIAB
ASSIGN PV
TAG
ALARM SAT VALUES
PASSWORD
SIMULATE
HART REV
BACK TO MENU
EXIT MENU
September 2020
HART7
2, 2, 5, 2, 3
93

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents