Configuration
September 2020
Table 2-6: Custom Alarm and Saturation Values
Level
Low
High
Failure mode alarm and saturation levels can be configured using a Field Communicator,
AMS Device Manager, and the LOI. The following limitations exist for custom levels:
•
Low alarm level must be less than the low saturation level
•
High alarm level must be higher than the high saturation level
•
Alarm and saturation levels must be separated by at least 0.1 mA (0.025 Vdc)
The configuration tool will provide an error message if the configuration rule is violated.
Note
Transmitters set to HART
digitally; saturation and alarm conditions will not affect the analog output. See also
Establishing multidrop
Configuring alarm and saturation levels using a Field
Communicator
From the HOME screen, enter the Fast Key sequence.
Device Dashboard Fast Keys
Configuring alarm and saturation levels with AMS Device
Manager
Procedure
1. Right select on the device, and select Configure.
2. Select Configure Alarm and Saturation Levels button.
3. Follow screen prompts to configure Alarm and Saturation Levels.
Configuring alarm and saturation levels using LOI
Refer to
26
4–20 mA (1–5 Vdc) saturation
3.70 mA– 3.90 mA (.90 –.95 V)
20.10 mA –22.90 mA (5.025 –5.725 V)
®
multidrop mode send all saturation and alarm information
communication.
Figure 2-13
for instructions to configure alarm and saturation levels.
4–20 mA (1–5 Vdc) alarm
3.60–3.80 mA (.90 –.95 V)
20.20 mA – 23.00 mA (5.05 –5.75 V)
2, 2, 2, 5
Rosemount 2051 Pressure Transmitter
Reference Manual
00809-0100-4107