Setting up CBCT Reconstruction Environment

1) Open APsetting.vxm and change GPUMODE = 4 to GPUMODE = 3 and save/exit the file.
VT-TECH-Smart Plus-001
About CT Reconstruction setting files
(Go to the following path = C:\VCapture\Acquisition\CBCT\WidePano\REC)
Files
APsetting.vxm
Reconstruction.vxm
Reconstruction_Bead.vxm
Reconstruction_GEO.vxm
Angle.bin,
RCP.bin,
RCPX_DBP.bin,
ROP_DBP.bin,
C:\VCaptureSW\Acquisition\CBCT \WidePano\Para folder always.
If there goes something wrong with Geometry calibration, delete the four files in the
current folder and recover the original files from the subfolder (./Base_para)
Make sure that there is no file other than these: Angle.bin, RCP.bin, Uoffset.bin,
Voffset.bin,
Angle_DBP.bin,
CriDetVal.bin in the C:\VCaptureSW\Acquisition\CBCT \WidePano\Para folder.