9) Type the mA value that meets the standard of the Background Level in the Green section. For
Woman, subtract 0.3mA from the mA in Man mode and for Child, subtract 0.6mA from the mA in
Man mode and type the value in the section.
10) Save and exit PwrParam_CBCT.ini.
VT-TECH-Smart Plus-001
ex) 5.0mA
ex) 4.7mA = 5.0mA - 0.3mA
ex) 4.4mA = 5.0mA - 0.6mA
Do not change the kVps.
Make sure that each FOV (FOV 0, FOV 1, FOV 2) has the same values.
By MAN_HARD=0800, 0500, '0800' means 80kVp and '0500' means 5.0mA.
75 / 188
Smart Plus (PHT-35LHS)
Technical Manual