Smart Plus (PHT-35LHS)
Technical Manual
2.
Aligning PANO Collimator
To optimize general image quality in Smart Plus (PHT-35LHS), the operator should perform the 1st
Collimator Alignment operation according to the following procedures.
4-axis Collimator
<1st Collimator Assy>
2.1. PANO Collimator Alignment Standard
-
- X-ray Condition: 60 kVp, 5 mA / Threshold Value: 70% (Adult), 90% (Child)
Top : 10~35 Pixels
Left/ Right : 2~4 Pixels
Bottom : 10~25 Pixels
VT-TECH-Smart Plus-001
118 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED