Performing Ceph Geometry Calibration - Vatech Smart Plus PHT-35LHS Technical Manual

Hide thumbs Also See for Smart Plus PHT-35LHS:
Table of Contents

Advertisement

Smart Plus (PHT-35LHS)
Technical Manual
2.

Performing CEPH Geometry Calibration

1) Follow the procedure below to perform the CEPH geometry calibration.
2) Run Console Software.
3) Click CEPH > High Resolution > PA > CONFIRM and wait until READY button is enabled.
4) Attach CEPH PHANTOM A'ssy and Quart Phantom as shown in the figure.
5) Click READY.
VT-TECH-Smart Plus-001
141 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED