Vatech Green 16 Installation Manual page 79

Table of Contents

Advertisement

5.1.2
Drilling the Floor
1.
Remove the aligning plate.
2.
Drill the floor holes (size: 10.5 x 30 mm) by using the concrete hammer drill.
3.
Remove the debris and clean the holes with the dust pump.
4.
Put the anchor bolts into the holes and attach them with the hammer. Verify that
the anchors are secured.
Anchor Bolt
Hammer Drill
Green 16/18 Installation Manual
5. Installing the Equipment: Wall Mount
5/16 x 60 w/ Spring and Flat Washers
2 pcs
(Part No. 34)
L = 200 mm (7.9")
71