Vatech Green 16 Installation Manual page 15

Table of Contents

Advertisement

Green 16/18 Installation Manual
so that the person can reach out to them easily in an
emergency.
The X-ray room must be shielded accordingly to the
radiation safety requirements to ensure the operator
and patient's safety. You must verify that the room
meets all applicable safety requirements.
Other devices must not be installed near the X-ray
equipment.
DO NOT install this equipment in an area that is
exposed to strong electromagnetic fields.
DO NOT install this equipment near the explosive
materials or in the explosion-prone area.
The electrical installation of this equipment must follow
all local code requirements for electro-medical systems:
IEC 60364-7-710:2002.
The UPS must be installed at the same time as the
equipment.
The equipment, PC, and all peripheral devices must be
well-grounded.
X-ray Radiation Warnings
Install this equipment in an approved location to ensure
the patient's and operator's safety.
Radiation Shielding (e.g., Stationary radiation shielding
barriers) must be installed with the equipment to protect
the operator from radiation exposure.
Follow all federal and local The X-ray system may
cause injury to the patient if improperly used. Obey all
federal and municipal standards regarding radiation
safety.
When the patient is taking an X-ray image, the operator
must stand behind a protective wall or take other
protective actions.
The operator must stay at least 2 m (6 feet) away from
the X-ray source when pressing the Exposure Switch
and observing the patient during the image acquisition.
The operator must provide the patient with protective
clothing before X-ray capturing.
Pregnant women must consult with a doctor before
being exposed to an X-ray.
1. Introduction
7