Vatech Green 16 Installation Manual page 14

Table of Contents

Advertisement

1. Introduction
6
Cautions for Installation
You must read and understand all instructions in this
manual before installing the equipment.
You must confirm that the system is installed according
to the instructions provided by this manual and perform
the appropriate procedures therein.
Before turning on the equipment, leave it a room
temperature for more than an hour
If the equipment has been stored at temperatures of
below 10℃ (50℉) for more than a couple of hours,
allow the equipment to reach room temperature before
applying mains voltage
Installation and related work must be performed by
people authorized by VATECH.
DO NOT connect any items or equipment that is not
part of this system: IEC60601-1-1 (3rd edition: 2005).
Any equipment not approved by VATECH must comply
with the applicable standards: IEC 60950-1 (2nd
edition: 2005) for IT equipment (Ex: PC) and IEC
60601-1 (3rd edition: 2005) for medical electrical
equipment.
All operators of this equipment must ensure that all
requirements outlined in IEC 60601-1-1 (3
2005): Safety Requirements for Medical Electrical
Equipment are fully met for the safety of patients,
operators, and the environment.
DO NOT touch-sensitive areas such as the sensor
when installing the equipment. The touch-sensitive
areas are mentioned in relevant sections.
The use of wireless phones may interfere with the
operation of this equipment.
Use an ESD (electrostatic sensitive device) wrist band
to connect it to a ground wire when installing the
equipment.
Touch a ground point to discharge static electricity
before handling PCB boards.
Important Notes for Installation Site
The PC monitor, emergency cut-off switch, and X-ray
Exposure Switch must be located close to the operator
edition:
rd
Green 16/18 Installation Manual