Determining the Height
1.
Determine the screw height using the following formula.
H
screw height
100 mm: the minimum desired distance between the ceiling and the top of the
equipment when the column is fully extended.
d = H
equipment without Base)
Ex) If
d = H
H
screw height
Green16/18 Installation Manual
=100 mm – d
- H
= H
– 2304 mm (Example hight: The height of the
ceiling
Max
ceiling
H
is 2200 mm, H
ceiling
- H
= 2200 mm – 2304 mm = -104 mm
ceiling
Max
= 100 mm - d = 100 mm + 104 mm = 204 mm
(If d is more than 100 mm, the column height limit is not necessary.)
value is calculated as follows:
screw height
Appendix