Selecting the Cold Start Menu Item

Use the following procedure to cold start the probe using the probe's Main Menu:
1. Connect a local terminal (or a PC emulation a terminal) to the probe. Refer
to "Probe Operation" on page 91 for information on connecting a local
terminal.
2. Press the CONFIG button once (on the back of the probe) to place the probe
into the configuration mode. Use a narrow, pointed object (like a pen) to
press the recessed CONFIG button. The probe then displays its Main Menu
on the terminal. Figure 5-1 on page 95 shows the probe's Main Menu.
3. Press 8 to execute a cold start and to exit the probe's Main Menu. The cold
start occurs immediately. The Activity and Fault LEDs are turned on during
a cold start. When the cold start completes, the Activity LED flashes to
indicate traffic (if present), the Fault LED turns off, and the ~Line On (or
Power) LED is on.
Probe Operation
Restarting the Probe
97