Instrument Reboot (Cold Start)

Installation Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
An instrument "cold start" routine is provided to reset the
instrument in the event of an unplanned hardware or firmware
event. A cold start reboots the instrument and restarts the
instrument using a default configuration file. Performing a cold
start erases existing configuration information. Current menu
settings will be lost and the instrument will power up in its
default state.
To perform a cold start
1 Switch the instrument off and wait a few seconds.
2 Switch the instrument on and as the instrument boots up,
look carefully at the display.
3 When the "Press the Menu key for startup options" text is
displayed, immediately press the <Menu> key.
4 Press 3 on the numeric keypad to select cold start. The unit
will then continue with the boot up process.
5 When the boot-up procedure is complete, the instrument
displays a dialog box with the message: "Instrument reset to
default settings." Press Select to close this window.