84
User Guide

Gas connection

Measurement gas inlet
Vacuum connection
Buffer gas inlet
Calibration gas inlet
Air inlet/Cal. gas overflow

Basic specifications

Warm-up time
Measuring ranges
Ambient temperature
Ambient pressure
Measurement gas
Measurement pressure
Vacuum
1
Vacuum >= 400 mbar
O

sensor

2
Sensor type
Measure gas temperature
Measure flow
P/N 320598-G
11/2019
EN
Hose Ø5/3 mm, sample hose l=3m
Hose Ø8/6 mm
Push-in, OD=3 mm
M5 female (normally plugged)
M5 female (normally fitted with filter)
Operational after 10 min.
(60 minutes to full specifications)
0-100% O
and optionally 0-100% CO
2
Operational:
0 to +45 °C, less than 95% RH,
non-condensing
Storage:
-10°C to +60°C, less than 95% RH,
non-condensing
Operational:900 hPa to 1050 hPa
Inert gasses (O
, CO
, N
, Ar, Air), less than 95% RH
2
2
2
Measure input:
Buffer versions - gas inlet:
< 100 mbar abs. @ flow of 5 l/min.
Zirconia
+0 to +35 °C
Measure input:
Sensor flow:
Buffer measurement:
Dansensor
MAP Check 3 Vacuum
®
2
1
0-1050 mbar
1.5-10 bar
Typically 5 l/min. ± 0.5 l/min.
75 ml/min. ± 5 ml/min.
75 ml/min. ± 5 ml/min.