42
User Guide
No.
85*
86
87*
88
89
90*
91*
P/N 320598-G
11/2019
Type
Message
Warning
Device temperature very low
Error
Device temperature too low
Warning
Device temperature very high
Error
Device temperature too high
Error
Zr O2 cold-junction temperature too low, shutting
down O2 sensor.
Warning
Zr O2 cold-junction temperature high
Warning
Ir CO2 sensor temperature low
EN
Dansensor
MAP Check 3 Vacuum
®
User action
Place unit in
temperature
above 0 °C
Allow unit to
heat up
Place unit in
temperature
above 0 °C
Allow unit to
heat up
Check cooling/
filters
If error persist,
contact service
Check cooling/
filters
If error persist,
contact service
Check ambient
temperature -
must be within
range
If problem
persists, contact
service
Check cooling/
filters
Check ambient
temperature -
must be within
range
If problem
persists, contact
service
Check ambient
temperature -
must be within
range
If problem
persists, contact
service