Alarm settings
Buffer alarm settings Setting of buffer measurement warning and alarm values.
Extra sampling time
EN
Setting of the sample gas measurement warning and alarm values.
Product alarm settings
%O
%CO
2
2
Off
Off
Off
Off
Off
Off
Off
Off
You have the following options for each alarm:
• Off
Disables all alarms
• On
Enables alarm for the selected input
- O
High alarm and Low alarm (%)
2
- O
High warning and Low warning (%)
2
- CO
High alarm and Low alarm (%)
2
- CO
High warning and Low warning (%)
2
If alarm settings are as shown above, O
and 0,7% are OK, measurements between 0,7 and 1,0% will result in
a high warning, an measurements above 1,0% will result in a high
alarm.
Any warning or alarm will result in a relay activated accordingly and
an indication in the display.
Buffer alarm settings
%O
%CO
bar
2
2
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Off
Buffer pressure is measured at every "EVAC" signal.
Any warning or alarm will result in a relay activated accordingly and
an indication in the display.
Prolongs the sampling period. If the gassing time is very short the
gas can spread through the system during the prolonged sampling
period.
Recommended setting is 1 sec.
Product alarm settings
%O
%CO
2
2
1,0
28,5
0,7
27,0
0,3
23,0
Ok
0,1
21,5
measurements between 0,3
2
Buffer alarm settings
%O
%CO
2
2
1,0
28,5
10,0
0,7
27,0
0,3
23,0
Ok
0,1
21,5
User Guide
Ok
bar
8,0
4,0
Ok
2,0
P/N 320598-G
11/2019
55