System leak test

This test is used to verify system tightness in sensor and flow parts. A leak is fatal for
measurements and the test should therefore be done periodically or in case of strange
readings.
Apply vacuum to the device and stop the packaging machine.
Enter Diagnostics and go to page 2 and select System leak test in the menu.
A popup window will remind you to stop machine and to ensure that vacuum is applied.
Press Yes to continue or No to cancel.
The test will start with a short stabilizing period then a 1 minute measurement is
performed.