Vibration isolation

Enclosure environment

38
Antaris II and Antaris MX Site and Safety Information
The table that follows lists the weights of rackmount components.
Component
Antaris II analyzer
Antaris MX analyzer
Rack drawer
(includes rails and fasteners)
The rollout drawer is designed to isolate the instrument from minor
vibrations caused by power supplies and small moving parts. Do not install
vacuum pumps, compressors, clean air or purge gas generators/driers or
other motor-driven devices in the same rack enclosure as the instrument
system. Motor-driven devices should be installed in a separate enclosure or
on a solid, level floor near the instrument enclosure. Vibrations from the
motors, compressors, and pumps can distort your spectral data.
To prevent damage to the instrument, you must provide cooling and
desiccating devices and maintain ambient temperature and humidity inside
the rack enclosure as follows.
Temperature — 15 °C and 35 °C (59 °F and 95 °F)
Humidity — 20 % to 80 % noncondensing
Cooling and desiccating devices must be available for the service engineer
at the time of installation. Damage to the instrument from improper
environmental controls is not covered under warranty.
Weight
60 kg
131 lb
26 kg
57 lb
15 kg
32 lb