Specifications

Power consumption

Purge gas

50
Antaris II and Antaris MX Site and Safety Information
Component
Antaris MX analyzer (including
modules)
Antaris II analyzer
(including modules)
External Heater with block holder*
Standard computer and monitor*
Integrated computer
Whatman purge gas generator
Whatman pure air dryer
Standard printer*
*Values shown are estimates. See the power specifications on the
rear panels or undersides of these units.
Locate the purge source within 10 ft of the instrument.
Item
Source line pressure
Purge pressure
Flow rate
Filtering
Dew point
90 W
110 W
100 W
460 W
*
70 W
10 W
1,000 W
200 W
Specification
20 psig min
150 psig max
1.4 atm min
10.2 atm max
10 psig min
20 psig max
0.7 atm min
1.4 atm max
15 scfh min
20 scfh max
7.1 l/m min
9.4 l/m max
10-micron
removes
filter
particulate
matter and oil
-70 °C
for best
(-94 °F) or
performance
below
Power Consumption
307 Btu/hr
375 Btu/hr
340 Btu/hr
1,570 Btu/hr
239 Btu/hr
34 Btu/hr
3,414 Btu/hr
683 Btu/hr