Rackmount Considerations; Tip Hazard - Thermo Scientific Antaris II Manual

Table of Contents

Advertisement

Location
Rackmount
The instrument, can be equipped for installation as roll-out components in
a standard 30 in (76.2 cm) rack enclosure that meets the EIA Rs-310C,
considerations
19 in (48.26 cm) rackmount standard. Six standard rack units (6U) are
required for the instrument.
6U
76.2 cm
48.26 cm
30 in
19 in

Tip hazard

Due to its weight, the instrument can pose a rack enclosure tip hazard
unless appropriate anti-tip extensions, feet, or anchors are in place. Use the
information that follows to have them designed by your rack enclosure
manufacturer or a qualified structural engineer prior to the installation of
the instrument. The structural engineer who designs your restraints will
need the weight information for the instrument components and any other
items installed in the rack enclosure.
Thermo Fisher Scientific
Antaris II and Antaris MX Site and Safety Information
37