Coolant gas

Purge gas generator

Thermo Fisher Scientific
Some rackmount enclosure require a flow of coolant gas. The following
specifications can be used to ensure that an unpurged instrument is
protected.
Item
Source line pressure
Purge pressure
Flow rate
Filtering
Dew point
The following illustrations show the dimensions of the purge gas
generators and the locations of holes that can be used for mounting the
generators on a wall. Models 75-45 and 75-45EU fit easily on top of a
bench or table. When not wall-mounted, Models 75-52 and 75-52EU
should be placed on the floor.
Specification
20 psig min
150 psig max
1.4 atm min
10.2 atm max
10 psig min
20 psig max
0.7 atm min
1.4 atm max
15 scfh min
20 scfh max
7.1 l/m min
9.4 l/m max
10-micron
removes
filter
particulate
matter and oil
-70 °C
for best
(-94 °F)
performance
or below
Antaris II and Antaris MX Site and Safety Information
Specifications
51