Location

Consider these points when planning the location of your system:
If possible, choose a site easily accessible by Thermo Fisher Scientific
personnel and have a telephone near the system. Should the system
require technical support or service, these measures will save time.
Be sure an AC power source is nearby. The instrument has a power
cord approximately 2.5 m (8 ft) long.
Be sure the installation surface is rigid, within 2 °° of level, and capable
of supporting twice the instrument weight.
While vibration will not damage the instrument, excessive vibration
may compromise spectral quality and measurements.
Make sure the system will fit in the work space you choose. Compare
the work space with the dimensions shown in the next section and in
the "Utility Requirements" chapter.
Leave adequate clearance around the system. Access covers must be
able to open fully for maintenance and service.
Place the system at a height that allows you to work comfortably.
Antaris II and Antaris MX Site and Safety Information
31