Location
Consider these points when planning the location of your system:
•
If possible, choose a site easily accessible by Thermo Fisher Scientific
personnel and have a telephone near the system. Should the system
require technical support or service, these measures will save time.
•
Be sure an AC power source is nearby. The instrument has a power
cord approximately 2.5 m (8 ft) long.
•
Be sure the installation surface is rigid, within 2 °° of level, and capable
of supporting twice the instrument weight.
•
While vibration will not damage the instrument, excessive vibration
may compromise spectral quality and measurements.
•
Make sure the system will fit in the work space you choose. Compare
the work space with the dimensions shown in the next section and in
the "Utility Requirements" chapter.
•
Leave adequate clearance around the system. Access covers must be
able to open fully for maintenance and service.
•
Place the system at a height that allows you to work comfortably.
Antaris II and Antaris MX Site and Safety Information
31