6 Preparing the product for use
6.1 Control programmes for 8E33=7-1, 8E34=7-1
The Digital Twin control unit features a digital and dual-channel control mode. The desired control
mode is selected with the integrated 13E189 function plug.
6.1.1 Digital mode programme (with function plug)
Control with 2 electrodes.
6.1.2 Dual-channel mode programme (without function plug)
Control with one electrode.
6.2 Control programmes for 8E33=9-1, 8E34=9-1
6.2.1 Programme 1: DMC plus
Proportional control with 2 electrodes for patients with two strong electrode signals
Depending on the strength of the electrode signal, the opening and closing speed can be varied
proportionally between 8 mm/s and 200 mm/s. The gripping force can also be varied proportion
ally between 10 N and 160 N.
Control
Closing
Opening
Gripping force
20
Function plug inserted
Opening/closing: Constant speed after the electrode signal
exceeds a threshold.
Function plug removed
Close: Slow, weak muscle signal
Open: Fast, strong muscle signal
The "757T13 MyoSelect" programming device is used to select the desired
control programmes.
The following control programmes are recommended:
Control programme 1 DMC plus: for patients with two strong electrode
•
signals
Control programme 3 VarioControl: for patients with one strong elec
•
trode signal
2 electrodes, proportional
Proportional to the signal strength on the "CLOSE" electrode
Proportional to the signal strength on the "OPEN" electrode
Proportional to the signal strength on the "CLOSE" electrode