2.3 Combination possibilities
This product can be combined with the following Ottobock components:
Power supply (rechargeable battery)
•
757B35=1
/
MyoEnergy Integral
•
757B20/757B21 EnergyPack
Battery chargers
The following battery chargers can be used
depending on the rechargeable battery:
•
757L20 charger (including 757L16* power
supply)
•
757L35 battery charger
3 Intended use
3.1 Indications for use
The product is intended exclusively for upper limb exoprosthetic fittings.
3.2 Conditions of use
The product was developed for everyday use and must not be used for unusual activities. These
unusual activities include, for example, sports which involve excessive strain and/or shocks to the
wrist joint (pushups, downhill racing, mountain biking, ...) or extreme sports (free climbing,
paragliding, etc.). Furthermore, the product should not be used to operate motor vehicles, heavy
equipment (e.g. construction machines), industrial machines or motor-driven equipment.
The product is intended exclusively for use on one patient. Use of the product by another person
is not approved by the manufacturer.
Permissible ambient conditions are described in the technical data (see page 25).
3.3 Indications
•
Amputation level below-elbow, above-elbow and shoulder disarticulation
•
For unilateral or bilateral amputation
•
Dysmelia of the forearm or upper arm
•
The user must be able to understand usage instructions and safety notices and put them into
practice.
•
The user must fulfil the physical and mental requirements for perceiving optical/acoustic sig
nals and/or mechanical vibrations
3.4 Contraindications
•
All conditions which contradict or go beyond the specifications listed in the section on
"Safety" and "Indications for use".
3.5 Qualification
The fitting of a patient with the product may only be carried out by O&P professionals who have
been authorised with the corresponding Ottobock training.
4 Safety
4.1 Explanation of warning symbols
Warning regarding possible serious risks of accident or injury.
WARNING
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
16
757B35=3
/
757B35=5
Lamination rings
•
10S1=50 lamination ring
•
9S110=50 lamination ring
Rotation
•
10S17 electric wrist rotator
•
13E205 MyoRotronic
Elbow
•
ErgoArm Hybrid plus: 12K44=*
•
ErgoArm Electronic plus: 12K50=*
•
DynamicArm: 12K100N=*/12K110N=*