Function
The 757T10 MyoSimulator has been designed to control MyoBock system components
during functional tests.
4.1
Simulation Possibilities
Simulates a maximum of two electrode signals.
Simulates a single movement (e.g., hand opening).
Simulates common movement cycles (e.g., hand closing-hand opening-hand
closing, etc.) along previously set levels
Co-contraction with set signal levels.
Safety co-contraction with set signal levels.
4 . 2
C o n n e c t i o n a n d S e r v i c e E l e m e n t s ( F i g . 1 , P o s . 1 - 1 2 )
Pos.1 COUNTER Shows the number of movement cycles. Shows "0" by a single
movement sequence.
Pos.2 OPEN
Set the desired level needed to open the hand here.
Pos.3 OPEN / CO-CONTRACTION
If the flip-switch (Fig.1, Pos.4) is set on normal use, then the set signal level
for this channel will remain selected for the duration of the test.
Pos.4 NORMAL / CO-CONTRACTION / SAFETY CO-CONTRACTION
Wählen Sie hier zwischen Normalbetrieb, Ko - Kontraktion und Sicherheits-
Ko-Kontraktion.
Pos.5 CLOSE
Set the desired level needed to close the hand here.
Pos.6 OPEN - CLOSE
Choose between "hand open" and "hand close" operating modes.
Pos.7 CLOSE / SAFETY CO-CONTRACTION
If the flip-switch (Fig.1, Pos.4) is set on normal use, then the set signal level
for this channel will remain selected for the duration of the test.
Pos.8 ADAPTER - Socket for connection to 757P40 connection cable.
Pos.9 MANUAL / AUTOMATIC Flip-switch for switching between single use (Man.)
and continuous use (Aut.).
Pos.10 T1- Alter the time period for the opening of the hand with the included scre-
wdriver.
Pos.11 Alter the time period for the pause between the opening and closing of the
hand with the included screwdriver.
Pos.12 Alter the time period for the closing of the hand with the included screwdriver.