Use
→ The navigation menu opens.
2) In the navigation menu, tap the entry "Functions".
3) Tap the entry "Mute mode".
4) Follow the on-screen instructions.

9.6 Deep sleep mode

INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see Page 38).
The Cockpit app can be used to place the knee joint into a deep sleep mode, in which power consumption is min­
imised. The knee joint offers no functionality in this mode. The safety mode damping values are activated.
It can be awakened from deep sleep mode with the Cockpit app or by connecting the battery charger. Waking
from deep sleep mode using the Cockpit app can take up to 30 seconds.
After ending deep sleep mode, the knee joint is in basic mode again.

9.6.1 Turning deep sleep mode on/off using the Cockpit app

Activating deep sleep mode
1) When the component is connected, tap the  symbol in the main menu.
→ The navigation menu opens.
2) In the navigation menu, tap the entry "Functions".
3) Tap the entry "Activate deep sleep mode".
4) Follow the on-screen instructions.
→ The activation of deep sleep mode is indicated by a short beep signal and a short vibration signal, assuming
that mute mode (silent mode) is not active.
Deactivating deep sleep mode
1) When deep sleep mode is active for the currently connected prosthesis, the Exit deep sleep mode button
automatically appears when the Cockpit app is started.
2) Tapping this button establishes a connection to the prosthesis and deactivates deep sleep mode.
INFORMATION: Establishing a connection in deep sleep mode can take up to 30 seconds.
If a prosthesis is in deep sleep mode but not connected to the Cockpit app, a connection to the prosthesis has to
be established (see Page 33).

9.7 OPG function (Optimised Physiological Gait)

INFORMATION
The "PreFlex" function can be activated/deactivated using the adjustment software.
All other parameters of the OPG function are always active and cannot be influenced.
The OPG function minimises the prosthesis wearer's prosthetic deviations from a harmonious gait pattern and pro­
motes more biomechanically correct walking. This function enables the following features:
PreFlex
PreFlex ensures the knee is in 4° of flexion at the end of swing phase in preparation for loading response. This
makes initiating stance phase flexion easier and forward movement is less restricted.
Adaptive yielding control
The knee joint has auto-adaptive stance and swing extension resistance. The stance flexion resistance experienced
by the user is dependent on the slope or incline when walking downhill. When walking on a ramp, adaptive yielding
control manages flexion depending on the angle of the ramp. The knee joint flexes slowly if the ramp is flat, and
flexes quickly if the ramp is steep.
Dynamic stability control (DSC)
DSC ensures the knee will not release stance resistance during biomechanically unstable static and dynamic con­
ditions. Constantly checking multiple parameters, DSC ensures the optimally timed decision for the knee to safely