Activating/Deactivating E-Mag Active - Otto Bock E-MAG Active Instructions For Use Manual

Hide thumbs Also See for E-MAG Active:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 11

3.2 Activating/Deactivating E-MAG Active

CAUTION
Risk of Falling. Missing signals during activation indicate a failure of the E-MAG Active. Remove the
battery immediately and reinsert it again. Should the error occur again, please contact your specialist
workshop immediately.
INFORMATION
If the knee joint is not unlocked for more than one hour, the control electronics automatically switch
to the stand-by mode. To reactivate the system, simply push the top button on the control electronics.
The activation is made by inserting the battery.
Every time the E-MAG Active is activated, the electronic control unit runs a short self-test that
emits the following signals:
Signal
System ready >
knee joint locked
System not ready >
Knee joint flexed
(open)
When the E-MAG Active is brought in the flexion, the warning signal in from of intermittent sounds
will be transmitted until the E-MAG Active has stretched/locked.
The deactivation is made by battery removal.
E-MAG Active
1
Visual feedback (LEDs)
All LEDs flash shortly
(white)
Either LEDs flashing perma-
nently (white and green)
Acoustic feedback
2
Sound/length
Rising
Intermittent
beep
Ottobock | 15

Advertisement

Table of Contents
loading

Table of Contents