Identifying The Cassette And Sending The Examinations To The Auto Qc² Software Station Using Nx - AGFA Auto QC2 User Manual

Table of Contents

Advertisement

Identifying the Cassette and Sending the Examinations
to the Auto QC² Software Station using NX
For a procedure on how to identify cassettes, refer to the User Manual of NX.
After the exposures are done and the images are sent to NX, the exposures
must be sent to the Auto QC² workstation. Refer to the User Manual of NX on
how to send images to the destination.
At the moment the images have been sent successfully to the Auto QC²
Software Station, the following conditions will be checked:
• The Auto QC² Software checks if the image received is a quality control
image.
• The Auto QC² Software checks the exposure type.
• The Auto QC² Software checks the rescale type of the exposure.
• The Auto QC² Software checks if an exposure with the same exposure SOP
Instance UID already exists.
Note:
Always check if Auto QC² Software is running before you send
images to the Auto QC² station.
Make sure to identify the cassette with the Auto QC² exposure type.
It is recommended to put the images in one study and name it Auto
QC².
Fast ID is not supported for the DICOM Digitizers (ADC Solo, ADC
Compact, ADC Compact Plus, CR 25, CR 35-X, CR 75, CR 85-X, DX-
G and DX-M) for scanning the cassettes with quality control
exposures, as high resolution images are required for quality
control procedures.
Fast ID for identifying the cassettes is supported for the DX-S and
CR 30-X Digitizer, in the sense that for these digitizer types this is
the only way to identify cassettes without ID Tablet.
Auto QC² | General Workflow | 53
2370C EN 20180330 1524