Temperature Subsystem - ADS TRITON+ Installation, Operation And Maintenance Manual

Table of Contents

Advertisement

Problem
Possible
Cause
Possible
Solution
Problem
Possible
Causes
Possible
Solutions

Temperature Subsystem

The following table contains general troubleshooting techniques for
the temperature subsystem corresponding to all sensors supporting
the TRITON+.
Note: Contact your ADS representative for further
diagnosis prior to replacing a sensor or monitor.
Maintenance and Troubleshooting
Pressure depth readings display a good pattern, but are
consistently incorrect.
Sensor physical offset may be incorrect.
Verify the sensor offset and reactivate with the correct
settings if necessary.
Individual PDEPTH_1 (_2) values consistently read 0
under wet pipe conditions.
Sensor may be faulty.
5-volt (blue) fuse on regulator may have blown.
Replace the Peak Combo Sensor.
Replace the 5-volt (blue) fuse on the regulator.
Caution: Do not connect a questionable sensor to
another monitor before verifying the integrity of the
sensor. A shorted sensor will blow the 5-volt (blue)
replacement fuse on the regulator in the monitor.
11-69