18. Reconnect the sensor cable(s) to the appropriate port(s)
(Channel 1 and/or Channel 2) on the monitor.
Inspecting, Cleaning, and Handling the
Sensors
Perform the following sensor inspections during regular site visits:
Verify that the installation ring, bands or Mounting Bar are
secure and clear of debris.
Carefully remove any debris that has accumulated on or around
the Long Range Depth Sensor.
Clean the face of the Submersion Sensor with a clean, moist
cloth as needed.
Clean the face of the Surface Combo Sensor/Ultrasonic Depth
Sensor gently with a soft brush, and wipe the sensor with a
clean, moist cloth.
Note: The crystals on the Surface Combo Sensor are
covered with a Teflon
wire brush or other abrasive, metallic device to clean these
sensors.
Verify that the face of the ParaDepth or ParaFlow sensor are
level side to side and front to back using Qstart
Refer to Chapter 4, Configuration and Activation for more
information.
Verify that the face of the Surface Combo Sensor/Ultrasonic
Depth Sensor is horizontally level with the flow. If necessary,
reposition the sensor according to the procedures in Chapter 5,
Sensor Installation and Connection.
Scrub the face of the AV|Gated, AV|Max and/or Peak Combo
sensor with a soft bristle brush.
®