(ExPAC), the Monitor Assembly shall be bonded to the
ExPAC earth with a conductor having a minimum cross-
sectional area of 4 mm

Other Conditions for Safe Use

The ADS Model 8000-FST-IM shall use only the 12-volt IS
Battery Pack (ADS p/n 8000-0043) for internal power.
Only ADS-approved sensors shall be connected to the ADS
Model 8000-FST-IM: ADS sensor types CSX series.
The only communication devices that may be connected to the
COMM + EXT PWR connector on the TRITON+ are the
USB Serial Interface (ADS p/n 8000-0337), the
ExPAC/XBUS/XIO power cable (ADS p/n 8000-378-10/25),
the Direct Connect Interface (ADS p/n 8000-0054), the
Inclinometer (ADS p/n 8000-0528), and the Sampler Cable
(ADS p/n 8000-0348).
Note: The Triton-Telog Comm Cable (ADS p/n 8000-
0054-01) also may be connected to the COMM + EXT
PWR connector on the TRITON+ to provide
communication between the Telog Ru-33 or Ru-35 and the
monitor. However, because the Ru-33 nor Ru-35 are not
certified for IS operation, an installation involving a
connection between these units will not be considered
intrinsically safe, unless the Telog unit is installed outside
the hazardous area.
Note: The Ayyeka Wavelet 4R Comm Cable (ADS p/n
8000-0054-03) also may be connected to the COMM +
EXT PWR connector on the TRITON+ to provide
communication between the Ayyeka Wavelet 4R monitor
and the TRITON+. However, because the Wavelet 4R is
not certified for IS operation, an installation involving a
connection between this unit will not be considered
intrinsically safe, unless the Ayyeka unit is installed
outside the hazardous area.