Running The Ground Wires - ADS TRITON+ Installation, Operation And Maintenance Manual

Table of Contents

Advertisement

6-22
ADS TRITON+ Manual
Positive and negative wires entering DC input terminal connector on ExPAC

Running the Ground Wires

Ground wires are required for all external power applications: one
running from the ExPAC to the earth ground and one running from
the ExPAC to the monitor. The procedure for installing these wires
varies slightly based on whether you are using the XIO, XBUS, or
another enclosure to house the ExPAC.
The XIO and XBUS both come from the factory with a short
ground wire running from the ExPAC to one of two ground
terminal blocks that are adjacent to the ExPAC. For these
applications, you must run a ground wire from one of the ground
terminal blocks to the existing earth ground outside the XIO/XBUS
and then run another ground wire from the other ground terminal
block to the ground connection facility on the TRITON+ monitor.
The ground connection facility represents the specific location on