Agilent Technologies 5530 Reference Manual page 38

Dynamic calibrator
Hide thumbs Also See for 5530:
Table of Contents

Advertisement

NOTE
1
Interferometer
assembly
2
Height adjuster and
post
3
Base
4
Laser head
Chapter 2 Linear Measurements
Mounting and Aligning the Optics on the Target Machine
Or, assemble and mount the retroreflector on the machine table as
shown in Figure 2-8.
If you plan on making a straightness or squareness measurement later,
position the top of the height adjuster even with the top of the post so you
can easily change optics. Then, use the adjustment knobs on the laser
head so the return beam enters the laser head's lower port.
1
Figure 2-7. Interferometer assembly table mounting
2-14
A
2
3
Measurements Reference Guide
4

Advertisement