Agilent Technologies 5530 Reference Manual page 254

Dynamic calibrator
Hide thumbs Also See for 5530:
Table of Contents

Advertisement

1
Surface plate
2
Measurement grid
3
Laser head
4
Nameplate
*
76-mm (3-inch) margin
between grid and
surface plate
Chapter 9 Flatness Measurements
Setting Up for the Measurement
2
1
3
*
Figure 9-3. Measurement setup when flatness mirrors are not used
If you will use a flatness mirror for one or more lines, leave at least a
102 mm (4 inch) border along one edge. This border allows sufficient room
for you to place the flatness mirrors.
The other sides require only a 51 to 76 mm (2 to 3 inch) border. Figure 9-4
shows an example of the measurement grid necessary when flatness
mirrors are used.