Chapter 4 Angular Measurements
Angular Position Measurements
Figure 4-32. Collect Data: ANGULAR screen
5
On the Collect Data: ANGULAR screen, select Reset Position.
This tells the Agilent 5530 calibrator that the current position of the
moveable optic (the machine's relative zero point) is the "zero" point from
which all angular positions will be measured.
If necessary, use the online help to complete the rest of the fields on this
screen.
6
Move the precision index table's base to the first measurement point.
7
Select Record.
If you are using the optional remote control unit, press the Record button.
Measurements Reference Guide
4-65