Chapter 8 Parallelism Measurements
Making Coplanar Parallelism Measurements
Continue moving the machine's moveable part to each point and recording
measurements until you are finished.
If at any point the Reset Laser message flashes, the laser beam path has
been broken. Move the reflector back to the zero position then select Reset
Position to reset the system.
After you record the last measurement, the Analyze Data:
PARALLELISM screen is displayed (Figure 8-9).
On the Analyze Data: PARALLELISM screen, select Next Axis.
You are returned to the Set Up Laser: PARALLELISM screen.
You are now ready to make the second axis measurement.
Making the second axis measurement
To make the second axis measurement, follow these steps:
Move the first axis moveable part to its start position.
On the Set Up Laser: PARALLELISM screen, do the following:
In the Measurement Axis box, select Coplanar.
In the Measurement Axis box, select W.
"W" tells the system this is the second axis measurement.
Repeat steps 3 through 8 in the last section, "Making the first axis
measurement."
Move the second axis moveable part. A pop-up window is displayed. The
parallelism value for the two axes is displayed in the Parallelism field.
Select Save Data to create or update a data file with the measurement
data. Refer to the Agilent 5530 Getting Started Guide and online help for
more information about using the Analyze Data: PARALLELISM screen.
Otherwise, you are finished making the measurement.
8-22