Chapter 6 Squareness Measurements in a Horizontal Plane
Making the First Axis Measurement
Continue moving the machine to each point and recording measurements
until you are finished.
If at any point the Reset Laser message flashes, the laser beam path has
been broken. Select Reset Laser to reset the system, then move the
retroreflector back to the last position measured before the beam was
broken.
After you record the last measurement, the Analyze Data: SQUARENESS
screen is displayed (Figure 6-11).
Figure 6-11. Analyze Data: SQUARENESS screen
On the Analyze Data: SQUARENESS screen, select Next Axis.
You are returned to the Set Up Laser: SQUARENESS screen. You are
now ready to mount and align optics for the second axis measurement.