NOTE
9
Chapter 8 Parallelism Measurements
Making Spindle Parallelism Measurements
Continue moving the machine's moveable part to each point and recording
measurements until you are finished.
If at any point the Reset Laser message flashes, the laser beam path has
been broken. Select Reset Position to reset the system.
After you record the last measurement, the Analyze Data:
PARALLELISM screen is displayed (Figure 8-9).
Figure 8-9. Analyze Data: PARALLELISM screen
On the Analyze Data: PARALLELISM screen, select Next Axis.
You are returned to the Set Up Laser: PARALLELISM screen.
You are now ready to make the second set of measurements.
8-18
Measurements Reference Guide