Advertisement

Chapter 4 Angular Measurements
Angular Position Measurements
NOTE
The example program, in Figure 4-18, does not include setting the relative
zero position, which you must do after you finish installing the
measurement hardware.
The measurement fixture clutch is designed to apply the desired clamping
force when the clutch control shaft has been pressed 1 mm in (down) from
its "neutral" position.
The precision index table requires a lift (up) of 2 mm from the
measurement fixture clutch's "neutral" position. The program example
assumes that you have set the Z-axis reference with that lift performed;
that is, the Z-axis reference position is with the top of the precision table
2 mm above its neutral position.
The example program in Figure 4-18 does not automatically cycle.
Measurements Reference Guide
4-33