Emerson Rosemount 3051SAL Reference Manual page 89

Table of Contents

Advertisement

Reference Manual
00809-0200-4802, Rev BA
Line pressure effect
(1)
Models
Ultra and Ultra for Flow
3051S_CD
Zero Error
± 0.025% URL per 1000 psi (69 bar)
Range 2 - 3
± 0.125% URL per 100 psi (6,89 bar)
Range 0
± 0.25% URL per 1000 psi (69 bar)
Range 1
Span Error
± 0.1% of reading per 1000 psi (69 bar)
± 0.15% of reading per 100 psi (6,89 bar)
Range 2 -3
± 0.4% of reading per 1000 psi (69 bar)
Range 0
Range 1
(1) For line pressures above 2000 psi (137,9 bar) and 3051S_CD Ranges 4-5, see the 3051S Reference Manual (document number 00809-0100-4801).
(2) Zero error can be calibrated out.
(3) Specifications for option code P0 are 2 times those shown above.
Mounting position effects
Models
3051S_C
3051S_L
3051S_T and 3051S_CA
Vibration effect
Less than ±0.1% of URL when tested per the requirements of IEC60770-1 field or pipeline with
high vibration level (10-60 Hz 0.21mm displacement peak amplitude / 60-2000 Hz 3g).
For Housing Style codes 1J, 1K, 1L, 2J, and 2M:
Less than ±0.1% of URL when tested per the requirements of IEC60770-1 field with general
application or pipeline with low vibration level (10-60 Hz 0.15mm displacement peak amplitude
/ 60-500 Hz 2g).
Specifications and Reference Data
(1)
(2)
(3)(2)(3)
Ultra, Ultra for Flow, and Classic
Zero shifts up to ±1.25 inH
With liquid level diaphragm in vertical plane, zero shift of up to 1 inH
with diaphragm in horizontal plane, zero shift of up to 5 inH
extension length on extended units; all zero shifts can be calibrated out; no span effect
Zero shifts to 2.5 inH
O (63,5 mmH20), which can be calibrated out; no span effect
2
Appendix A: Specifications and reference data
Classic
Zero Error
± 0.05% URL per 1000 psi (69 bar)
± 0.125% URL per 100 psi (6,89 bar)
± 0.25% URL per 1000 psi (69 bar)
Span Error
± 0.1% of reading per 1000 psi (69 bar)
± 0.15% of reading per 100 psi (6,89 bar)
± 0.4% of reading per 1000 psi (69 bar)
O (3,11 mbar), which can be calibrated out; no span effect
2
(2)
(3)
O (25,4 mmH
2
O (127 mmH
O) plus
2
2
May 2013
O);
2
81

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 3051shpRosemount 3051s

Table of Contents