Reference Manual
00809-0200-4802, Rev BA
B.1.5
Hazardous locations certifications
North American Certifications
Factory Mutual (FM) Approvals
I5
CSA - Canadian Standards Association
Process sealing
All CSA hazardous approved transmitters are certified per ANSI/ISA 12.27.01-2003.
I6
European Certifications
I1
Special conditions for safe use (x)
The surface resistivity of the antenna is greater than one gigaohm. To avoid electrostatic charge
build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
1180
Product Certifications
FM Intrinsically Safe, Non-Incendive, and Dust Ignition-proof.
Intrinsically Safe for Class I/II/III, Division 1,
Groups A, B, C, D, E, F, and G.
Zone Marking: Class I, Zone 0, AEx ia llC
Temperature Codes T4 (T
Non-Incendive for Class I, Division 2, Groups A, B, C, and D.
Dust Ignition-proof for Class II/III, Division 1,
Groups E, F, and G.
Ambient temperature limits: -50 to 85° C
For use with Rosemount SmartPower options
00753-9220-0001 only.
Enclosure Type 4X / IP66
CSA Intrinsically Safe
Intrinsically Safe for Class I, Division 1, Groups A, B, C, and D.
Temp Code T3C
Enclosure Type 4X / IP66
For use with Rosemount SmartPower options
00753-9220-0001 only.
ATEX Intrinsic Safety
Certificate No.: BAS01ATEX1303X
Ex ia IIC T4 (T
= -60 °C to 70 °C)
a
IP66
For use with Rosemount SmartPower options
00753-9220-0001 only.
= -50 to 70° C)
amb
II 1G
Appendix B: Product Certifications
May 2013
141