(8) Material specified is cast as follows: CF-8M is the cast version of 316 SST, CF-3M is the cast version of 316L SST. For housing, material is aluminum with polyurethane paint.
(9) Available with output code A only. Not available with approvals. See Rosemount 753R Product Data Sheet, 00813-0100-4379, to specify Web-Based Monitoring Indicator.
Does not integrate into plant host systems.
(10) Available with output code A only. Available approvals are FM Intrinsically Safe, Division 2 (option code I5), ATEX Intrinsic Safety (option code I1), or IECEx Intrinsic Safety
(option code I7). Contact an Emerson Process Management representative for additional information.
(11) Not available with Wireless Operating Frequency and Protocol option codes 1 or 2.
(12) Not available with Wireless Operating Frequency and Protocol option codes 1 or 2 or Housing code 01. This option is only available with differential pressure ranges 2 and 3,
and silicone fill fluid.
(13) Thermowell material is the same as the body material.
(14) Applies to assembled flowmeter only, mounting not tested.
(15) Not available for bore sizes 0010, 0014, 0020, or 0034.
(16) Isolation and Instrument valves not included in Traceability Certification.
(17) Not available with DIN Process Connection codes D1, D2, or D3.
(18) Materials of Construction comply with metallurgical requirements within NACE MR0175/ISO for sour oil field production environments. Environmental limits apply to certain
materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments.
(19) Only available with Measurement Type D.
(20) Not available with Output Protocol code F or Housing code 01.
(21) Not available with output code F. Not available with housing code 01 or 7J.
(22) Only available with Measurement Types 1-7.
(23) Not available with M20 or G ½ conduit entry size.
(24) Consult an Emerson Process Management representative for availability.
(25) Not available with Housing code 01 or 7J.
(26) Not available with output code F, Housing code 01, option code DA1, or option code QT.
(27) Cable supplied is Belden 3084A, rated for ambient temperatures up to 167°F (75°C).
(28) Not available with Housing code 00, 01, 5A or 7J.
(29) Available with Output Protocol code A and PlantWeb housing only.
(30) Requires PlantWeb housing and output code F.
(31) Requires PlantWeb housing and output code A. Includes Hardware Adjustments as standard.
(32) Requires Rosemount Engineering Assistant version 5.5.1 to configure.
(33) Not available with Housing code 00, 01, 5A, or 7J. Available with Intrinsically Safe approvals only. For FM Intrinsically Safe, Division 2 (option code I5) or
FM FISCO Intrinsically Safe (option code IE), install in accordance with Rosemount drawing 03151-1009 to maintain outdoor rating (NEMA 4X and IP66).